1. Topic

  How to measure PM10?

2. Introduction

   

PM10 is the fraction of airborne (suspended) particulate matter which contains particles of diameter1 less than 10 μm. Airborne particles have a large range of diameters, from nano-particles and ultrafine particles (diameters less then 0.1 μm) to the very large particles with diameters up towards 100 μm. Such large particles are suspended in air only a short time after their release (they fall out to the ground because of their mass. Even larger particles are not considered as airborne particles at all). PM10 includes all particles, of different sizes and types, which are relevant for health effects. Thus, PM10 is the fraction which is regulated presently by the EU Air Quality (AQ) Directives. Sub-fraction of PM10 of smaller diameters, notably PM2.5 (diameter less than 2.5 μm), or even smaller size fractions (e.g. PM0.1, less than 0.1 μm) are presently considered to be more relevant for health effects than particles above 2.5 μm. PM2.5 is, to some extent, regulated by the AQ Directives already, and more attention will be given to PM2.5, and possibly even smaller fractions, in the future.

The 1st AQ Daughter Directive prescribes the extent to which Member States should measure PM10 and PM2.5, as well as the methods to be used.


3. Discussion

   

About PM mass monitoring methods

There are several methods for measuring the concentration of airborne PM mass. Methods have been developed over decades, and they have been improved in their sophistication. The basic technique is to pull a known volume of air through a filter, and weigh the filter under controlled conditions before and after the sampling. An important part of such type of PM instrument is the air/particle intake device, which separates the particles that should be measured (i.e. particles of diameter less than 10 μm, or particles less than 2.5 um) from the larger ones. Different types and makes of inlet separators have different separation performance.

This method is the basis for the Reference method for PM10 measurement described in the Directive, which actually specifies the actual instruments (products) which are accepted. The reason for the need to specify the instrument makes in the reference method is that the combination of air intake and filtering unit design defines the PM sample which is actually collected for weighing, and for various instrument designs/makes, the PM sample will differ. Thus, to accept all instrument makes of a certain type as reference method without testing, would introduce an uncontrollable variation results from reference PM instruments. The pragmatic solution was to select a certain small number of instruments as reference samplers. Other filtering type methods can also be used, but the user has to show that they are equivalent to the reference instruments, within certain prescribed level of accuracy.

Automatic instruments which are now available measure PM concentrations continuously, providing hourly values, while filtering methods generally provide 24-hour averages. These are used extensively in monitoring networks in Europe. Through experience and research is it clear that such instruments most often give significantly lower concentrations than filtering methods. Also the automatic methods are based on collection of the particles on filters, and various sensing techniques determine indirectly the mass of the PM. The filters are kept at elevated temperatures in these instruments to avoid problems related to humidity, and this elevated temperature results in loss of semi-volatile PM mass (mostly ammonium nitrate and organic compounds, of which the nitrate is most often the dominating one in terms of mass losses). Automatic monitors also have to be tested for equivalency with the reference method, to be accepted for monitoring under the AQ Directives. The most often used automatic monitoring methods are the beta attenuation method and the TEOM method. These and other methods are described in the 1st PM10 Position Paper (see web link section below).

Methods in relation to the assessment and information requirements in the 1st AQ Directive

The AQ Limit values for PM10 relates to daily (24-hour) averages and annual averages. Such statistics are also required from the PM2.5 measurements that the Member States have to set up. To assess these averages by measurements, filter methods giving 24-hour averages have sufficiently good time resolution. However, if there is a need to assess which sources give the main contributions to the measured values (This is needed when action plans for pollution reduction must be worked out), a better time resolution is generally needed, such as hourly data. Also, the Directive requires that information about PM10 concentrations shall be made available to the public “as soon as possible”, especially when limit values are exceeded. This requires the use of automatic methods, since with gravimetric filter methods data are available, at the earliest, several days after the sampling.

Reference methods

PM10

The following types of instruments, utilising a plane filter of certain specifications as the particle collection medium, can be assessed as being reference instruments for PM10 measurements in Europe:

ˇ medium-volume filter samplers;

ˇ high-volume filter samplers;

ˇ very-high-volume filter samplers (called WRAC-type samplers).

They all have to be equipped with suitable air intakes that separate the PM10 fraction from the larger particles, with effectiveness according to set specifications.

This is described in CEN Standard EN 12341:1998 – “Determination of the PM10 fraction of suspended particulate matter – Reference method and field test procedure to demonstrate reference equivalence of measurement methods”.

PM 2.5

A decision on reference methods for PM2.5 determination in Europe is under development (CEN standard prEN 14907, Under approval, as of December 2004).

Determination of equivalence of other instruments and methods

The procedure for determining whether a candidate method is equivalent to the PM10 reference method is described in the CEN Standard EN 12341.

The concept is that to use another instrument or method, a correction factor (CF) must be determined, so that when the data measured by a non-reference method is corrected by the CF, the resulting value is equivalent with what the reference method would give. The accepted level of uncertainty in the measurements as compared to the Reference method is prescribed in the standard. It is necessary to show that the CF is valid for area and PM10 level and composition (city, region, country) where the monitoring in carried out.

A report on the use of Correction Factors for PM10 determinations with various instruments in Europe has been worked out by the European Topic Centre on AQ and Climate Change (ETC/ACC) http://air-climate.eionet.eu.int/ of the European Environment Agency (EEA): “Correction factors and PM10 measurements”, available as a final draft report (as of December 2004).

The similar procedures for determining equivalence of PM 2.5 methods is described in the CEN standard prEN 14907.

Quality control of monitoring data

Annex VIII of the 1st Daughter Directive (see web link below) specifies the required accuracy of measurement data, as well as the minimum data capture (parts of the year the measurement data must be available from a monitoring station). These requirements are laid down as a guide to what quality-control and –assurance (QA/QC) programmes that monitoring network operators need to follow to comply with the quality objectives.

QA/QC procedures are dealt with in the Topic How to secure the quality of the monitoring data? Quality assurance (QA) and Quality Control (QC) systems and procedures..

Minimum number and location of PM10 monitoring stations in a Member State, according to the 1st AQ Daughter Directive

Annex VII of the 1st Daughter Directive (see web link below) state the requirements to what type of areas (zones) should be monitored for PM10, and the minimum number of sampling stations as a function of the population in an agglomeration or zone. In zones where the upper assessment threshold is exceeded and there should be more than 1 station, there should be at least one urban background and one traffic-oriented station.

The determination of which zones shall have PM10 monitoring stations is generally done through the activities of the “Preliminary Assessment” that the Member State shall carry out, in accordance with the Directive on ambient air quality assessment and management (the Framework Directive) (96/62/EC) (see web link below).

For PM2.5, each EU Member State should operate a number of monitoring stations, the number to be determined by each State so that PM2.5 concentrations representative for the variation of PM 2.5 within the State can be determined.

Location of sampling points, and exposure related monitoring

Annex VI of the 1st Daughter Directive presents considerations that shall apply regarding the selection of fixed point measurements. The annex describes macroscale siting and microscale siting considerations, both for health- and ecosystems-protection related stations.

Stations directed at human health protection should be located such that they provide data on areas representing typical or maximum concentrations that the population in the area is likely to be exposed to, directly or indirectly. This exposure should occur for a period which is a significant part of the averaging period of the limit value. Thus, it is prescribed that the monitoring should be exposure related.

This means simply that health-protection directed stations should be located where people are living or frequenting, and that the period that they are likely to be exposed in the area is comparable to the averaging time of the limit value. For PM10, the averaging times of the limit values are either 24 hours, or one year. So, monitoring stations for PM10 should be positioned either in residential areas where people are potentially exposed throughout the whole year or indeed during any 24-hour period, or they should in “hot-spot” areas where people are likely to be exposed over a significant part of any 24-hour period. This could be residences very close to roads with high traffic, or in areas close to industrial sources with large PM emissions.

This also means that areas where exposure is likely only over shorter periods, such a one or a few hours, should not be considered for monitoring of PM10.

Guidance on selection of monitoring method(s)

ˇ The selection of which method(s) to use for PM monitoring, in response to the requirements of the Directive, is influenced by, and is in practice a trade-off, between several factors: compliance with reference method prescription, need for fast availability of data, and costs (both for purchase, maintenance and system operation).

ˇ Reference method requirements: would be easiest to just use the reference instruments.

ˇ Need for fast information on PM levels (faster than several days): automatic method is needed. This requires that correction factor is determined for each area in question and

ˇ Minimisation of total costs: need to check prices. Automatic instruments are not always more costly than gravimetric filter samplers. Automatic instruments require fairly costly CF determination. Operating costs should be considered: manpower costs for field and laboratory operations; calibration needs and laboratory resources, equipment maintenance.


4. Recommendation / Conclusion

   

High quality PM10 and PM2.5 monitoring requires more resources and experience than monitoring of the typical gaseous pollutants such as SO2 and NO2. Reference laboratories in most countries have acquired much experience with PM10 monitoring and can give valuable advice to local authorities. It is highly recommended to seek such advice, even before purchasing instruments, if such experience is not available locally. The determination of locally valid correction factors (CF) for non-reference instruments is an important part of PM10 monitoring, if the resulting data is to be accepted by the EC. Advice should be sought with experienced institutions on CFs already determined for similar areas, and on the need to do additional CF determinations.


5. Examples / Further Reading

   

PM10 Monitoring and Intercomparison with the Reference Sampler in Helsinki

http://www.fmi.fi/kuvat/FINAL_PM_report_30_1_2004.pdf

Intercomparison between the TEOM analyser and the European reference sampler for the determination of PM10 concentrations

http://www.airquality.co.uk/archive/reports/cat13/0406301532_Intercomparison_report_FINAL.pdf


6. Additional Documents / Web Links

   

ˇ Air Quality Framework Directive 96/62/EC, OJ L 296, November 21, 1996 (http://europa.eu.int/smartapi/cgi/sga_doc?smartapi!celexapi!prod!CELEXnumdoc&lg=en&numdoc=31996L0062&model=guichett (choose .pdf file)

ˇ First daughter Directive 1999/30/EC on SO2, NO2, PM10 and Lead: http://europa.eu.int/smartapi/cgi/sga_doc?smartapi!celexapi!prod!CELEXnumdoc&lg=en&numdoc=31999L0030&model=guichett (choose .pdf file)

ˇ EU 1st Position Paper on PM10: http://www.europa.eu.int/comm/environment/air/pdf/pp_pm.pdf

ˇ EU CAFÉ 2nd Position Paper on Particulate Matter (2004): this is available as a Final Draft report. Contact the CAFE Secretariat, DG Environment, Brussels.


Last Updated


 

25th January 2005

Back