Running air quality monitoring stations is more then just
watching the numbers rolling in. To produce results of
known and sufficient quality there is a whole range of tasks to be performed
such as periodic status checking, maintenance, calibrations, data evaluation
and so on. Failing to perform all or some of these tasks will reduce the data
quality. To ensure unified operation both within a monitoring network and
across several networks a documented quality system is necessary. All
operations must be described in written procedures and documented for later
reference. Only then will it be possible to assess the quality of the measured
data as required in the EU Air Quality Framework Directive.
Based on the above and article 3 of the AQ Framework
directive which calls for the designation of bodies responsible for ensuring
accuracy of measurements etc., the Norwegian Pollution Control Agency appointed
NILU as National Reference Laboratory for Air Quality in Norway (NRL) in 2001. NRL’s first task was to introduce a common quality system
for all network operators in Norway.
The
Quality Manual was written for non-experts. It aims at guiding the operators
towards a unified way of operating their measurement networks by explaining in
detail how maintenance, calibrations, data validation, etc. is performed and
documented.
The main
documentation at a site is:
- Standard Operations Procedure
(SOP) for each instrument at the site
- A format for each SOP to
document the procedure
- Performance Acceptance
Criteria specific to the instrumentation at the site
The
results from periodic testing of instruments, typically once a week, e.g.
zero/span checks on gas monitors, are compared to the Performance Acceptance
Criteria. If the results fall outside the prescribed limits, certain actions
have to be taken, e.g. recalibration, troubleshooting or service of the
instrument. By using common action criteria across and between measurement
networks, it is more probable that the operators will evaluate test results
equally.
A typical
weekly maintenance procedure for a gas monitor would be:
- Record time and date of
arrival to the station, serial number of instrument and working gas
standard and working gas concentration.
- Record selected status
parameters, such as sample flow rate, sample temperature, reaction chamber
temperature, light intensity.
- Compare status parameters
with last few weeks of status parameters to detect trends, e.g. falling
light intensity in an ozone monitor. The objective is to change
consumables and spare parts before they reduce the quality of the data or
brake down.
- Disconnect the inlet tube
from the inlet and connect it to the zero/span check unit.
- Test the instrument by
feeding zero and span gas to it.
- Compare the results with the
results from last few weeks of zero/span checks to detect trends. Normally
there would be a linear decrease in the response. A sudden drop may
indicate a problem.
- Compare the results with the
performance acceptance criteria and perform necessary actions if the test
results fall outside the performance limits. The instrument response is
changed only if it is outside the action limits.
- Perform regular maintenance
as required, such as changing inlet filter.
- Remember to reconnect the
inlet tube to the inlet!
- Record time of end of
operation.
- Sign the maintenance form.
- Record the visit in the visit
log.
Once the
operator is back at the office he/she immediately calculates scale factors
(slope and offset) based on the results from the zero/span check. The scale factors
are entered into the data acquisition system and used by the system to
mathematically correct the acquired data. Data is collected every hour from all
stations, scaled and transferred directly for display on the internet web page
(in this case: http://www.luftkvalitet.info/
).
Even if
the instruments are maintained in a proper order they may break down. In order
to detect malfunctions as soon as possible, validation of collected data is
required. Continuous display of data on the internet requires some automatic
validation, re. e.g. spikes, too high values, too
negative values. In addition a manual data validation is performed as well to
cover other kinds of invalid data.
Manual
data validation is performed as follows:
- Every day the operator goes
onto the internet and looks at charts from his/her stations. The operator
looks for indications of instrument malfunctions, such as constant levels,
spikes and negative values.
- If any measurement data looks
suspicious the operator will connect to the station and run a check on the
instrument and inspect on-line or one-minute averages for details.
- Every week the operator
enters the scale factors into the system.
- After the end of the month
the last month of data is evaluated. The operator looks for trends in
measurement data and scale factors, invalid measurement data, e.g. from
zero/span checks and periods of instrument malfunctions are flagged.
- The monthly data is finally
approved and transferred to the central database. They are now ready for
use.
Calibrating
the instruments and the working gas standard at the site is done once every
three months, by bringing a travelling standard to the site.
The
instruments are brought to the lab at NILU once a year for major service,
adjustments and calibration. The calibration includes a test for linearity
(dynamic calibration).
To test
the actual performance of the networks, NRL performs an annual audit. The audit
includes a performance audit and a system audit. The performance audit is
conducted in every measurement network at a minimum of one measurement site
selected at random. It tests the actual performance of an instrument by using
an independent calibration standard brought to the site by NRL. The system
audit is performed to document if the network operator is operating in
accordance with the quality system and that the documentation and archives are
in proper order.
NRL
maintains the national reference gas standards and provides the networks with
trace-ability to these standards. To assure the quality of the standards and
other calibration equipment, NRL participates in international
inter-comparisons.
Attached
is an example of a SOP for weekly maintenance of a SO2 monitor.
ANNEX: Extract from NILU AQ Data Quality Manual
(see pages below)
Purpose
of SOP
To maintain the gas monitor in a good order.
The monitor needs
periodical maintenance and status checking to provide reliable results. The
maintenance and status checking includes changing of consumables, cleaning,
performance check, manual zero/ span check, leak checking and registration of
status parameters.
Applicability
and description of equipment
This SOP applies to the maintenance of gas monitors.
At the station the monitor is connected to the sample air
intake system. During zero/span checks the monitor is connected to an external
zero/ span check unit. The monitor is connected to the data logger by
electrical wires.
Responsibilities
Personnel performing routine
maintenance on gas monitors will be thoroughly knowledgeable of the contents of
this SOP and will comply with its requirements when performing routine
maintenance on gas monitors.
Instrumentation
This SOP assumes the following instrumentation:
·
TEI 145 permeation tube zero/ span check
calibrator
·
TEI model 42C SO2 monitor
Documentation
This SOP is documented in the form:
Routine maintenance, TEI model
43C SO2 monitor
Maintenance
procedure
1.
Enter todays date and the
time you entered the station in the Date
and Start time fields respectively.
2.
On the monitor press the RUN pushbutton to enter the continuous display. Enter the displayed
SO2 concentration in the Ambient -
Monitor - SO2 field.
3.
On the monitor press the MENU pushbutton to enter the monitor menu and choose ALARM. Enter the INTERNAL TEMP, CHAMBER TEMP,
PRESSURE, FLOW, INTENSITY, LAMP VOLTAGE and SO2 CONC status values in their respective fields in the Alarm section. Press RUN to return to the Run screen.
Manual
zero/ span check
A manual zero/ span check is
performed to validate the monitor response.
Registration of general
information:
1.
Enter the span gas cylinder pressure in the Z/S
check monitor – Wrk. std. – Pressure field.
2.
Enter the gas cylinder’s certified concentration in the
Cert. conc. field.
3.
On the monitor press the MENU pushbutton to enter the monitor menu and select RANGE using the arrow buttons. Press ENTER to enter the RANGE menu. Enter the RANGE
values in the Z/S check monitor -
Response - Range field in the form. Press RUN to return to the Run screen.
Zero check:
1.
Disconnect the inlet tube from the inlet and connect it
to the zero air generator.
2.
Adjust the flow to 1 l/min.
3.
Let the monitor sample zero air for 20 minutes. Enter
the 1 Minute Average value from the datalogger in the Z/S check Monitor - Response - Zero
field in the form.
Span check using a gas cylinder connected directly to the
monitor:
1.
Disconnect the inlet tube from the zero air generator and connect it to the span gas cyinder.
2.
Adjust the flow to 1 l/min.
3.
Let the monitor sample span gas for 20 minutes. Enter
the 1 Minute Average value from the datalogger in the Z/S check monitor - Response - Span
field in the form.
4.
Reconnect the inlet tube to the inlet.
Do not forget:
1.
Tick off the items in their respective fields in the NB! section in
the form.
2.
The monitor is sampling air from the intake.
3.
Gas cylinder regulators are closed.
4.
The zero/ span check unit and solenoids are not
activated.
5.
Sign the form
Routine
maintenance. TEI model 43C SO2 monitor
|
|
|
|
|
Station
name
|
|
Station Id.
|
Serial
numbers
|
|
|
|
|
|
|
AQ-
|
Monitor
|
|
Z/S
unit
|
|
Wrk. std.
|
|
|
|
|
|
|
|
|
|
|
Date
|
|
|
|
|
|
|
|
|
Start
time/ Stop time
|
|
|
|
|
|
|
Ambient
|
Monitor
|
SO2
|
|
|
|
|
|
|
|
|
Internal
|
|
|
|
|
|
|
|
Temp.
|
Chamber
|
|
|
|
|
|
|
|
|
Pressure
|
|
|
|
|
|
|
Alarms
|
Sample
|
Flow
|
|
|
|
|
|
|
|
|
Intensity
|
|
|
|
|
|
|
|
Lamp
|
Volt
|
|
|
|
|
|
|
|
Conc.
|
SO2
|
|
|
|
|
|
|
|
|
Pressure
|
|
|
|
|
|
|
|
Wrk. std.
|
Cert.
conc.
|
|
|
|
|
|
|
Z/S
check
|
|
Range
|
|
|
|
|
|
|
monitor
|
Response
|
Zero
|
|
|
|
|
|
|
|
|
Span
|
|
|
|
|
|
|
|
Intake
tube in place
|
|
|
|
|
|
|
NB!
|
Cylinder
valves closed
|
|
|
|
|
|
|
|
Logger
monitor off
|
|
|
|
|
|
|
Comments
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Signature
|
|
|
|
|
|
|
|
|
|